Standard Operating Procedure
Thermo Nicolet™-FTIR ATR

Infrared Spectroscopy
.

FTIR Setup and Operation Instructions
pdf version


Chemistry Department
Miramar College

FTIR Instrument


 

Precautions

• Remove gloves when operating the computer keyboard and mouse

• Remove your sample from the ATR, by using a Kimwipe® that is damped with acetone.

• Don’t forget to clean the plunger, if it made contact with your sample.

• DONOT squirt acetone directly on the platform EVER!!

 


 

Quick Overview

I. Running a sample  

1. Turn on the computer. The spectrometer should be on the entire time line  

2. Launch EZ OMNIC on the desktop.  

3. Make sure the bench Status is √ on the top right corner of the window. If it is not, please inform your instructor or the lab tech.  

4. Make sure that there isn’t anything left on the ZnSe crystal.  

5. The Experiment: window should also read {Smart Performer}.  

6. Do not change any of the parameters without instructor’s permission.

7. Click on Col Bkg to take a background of air. When the confirmation window pops up, click YES. The number of scan on the left bottom corner should increase as the spectrometer is recording your background.  

8. When the confirmation window pops up to ask to add Window 1, click NO.  

9. Place your sample on the ZnSe crystal- Liquids -place a drop of sample on the ZnSe crystal with a medicine dropper. Solids – place a small neat sample on the ZnSe crystal with a spatula. Dial the knob of the ATR until the pressure reads 12.  

10. Click on Col Smp , Collect Sample Window pops up, type in “sample name” and enter, confirmation window pops up, click YES. The number of scan on the left bottom corner should increase as the spectrometer is recording your sample.  

11. When the confirmation window pops up to add Window 1, click YES.    


II. Working up the Data
 

12 To have peaks labeled, Click Find Pks. Adjust the threshold by clicking on the window with the left mouse button. When peaks are selected, Click on “Replace”.  

13 If for some reason some of the peaks are not labeled, extra peaks can be manually labeled by clicking on the “T” on the bottom left corner. Use the “Text” too to select and or write a label or description.  

14. If satisfied with the information on the computer displaced spectrum, then WYSIWYG can be printed by clicking on Print icon and then print again in the print window.


 

Sign in Log book and Turn on computer

 

Activate EZ OMNIC

 

Check Bench Status

 

NO sample in ZnSe crystal

 

Check Smart Performer status

 

Collect Background

 

Scan Background

 

Do not add window for background scan

 

Liquid Sample on ZnSe crystal

 

Solid sample in ZnSe crystal

 

Collect spectrum for sample

 

Spectrum for sample

 

Adding Sample Name and Open New Window

 

Working up data, identifying peak labels

 

Label peaks

 

Print Spectrum

 

Clean up

 

Shut down

.