Spectronic 20™

<HR width=600>
PreviousNext
Description
Preparing to Use a Spectrometer
About the Meter
Reading the %T from the Meter
Setting the Wavelength
Setting Maximum Absorbance / 0% Transmittance
Setting Zero Absorbance / 100% Transmittance
Preparing the Cuvet and Placing the Sample in the Spectrometer
Measuring the %Transmittance
Additional Topics
Colors
Recording an Absorption Spectrum with a Spectronic 20™
Self Check Exercises 1
Self Check Exercises 2
Related Modules
Spectrometer, Scanning
Spectroscope, Hand-held
PreviousNext

Setting Maximum Absorbance / 0% Transmittance

 

Video. Setting the maximum absorbance.


Video. Self Check video.

The maximum absorbance is set with the sample compartment empty. The maximum absorbance is measured when no light is transmitted to the phototube detector. It is set by adjusting the power switch/zero control knob until the instrument shows a reading of 0 %T (or infinite absorbance).

check The Spectronic 20™ in the Self Check video is being set for maximum absorbance. Would any subsequent measurements be accurate? Why or why not? Answer

Accurate absorbance measurements cannot be obtained. The maximum absorbance has been set off the scale of the spectrophotometer. Instead the maximum absorbance should be set to 0%T before making other measurements.